Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Omega HH82: (1 year) CAL VER /5520 DATE: 06-Jul-98 AUTHOR: Fluke Corporation REVISION: $Revision: 1.2 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 18 NUMBER OF LINES: 98 CONFIGURATION: Fluke 5520A ============================================================================= # # Source: # Omega HH81 & HH82 Instruction Manual # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F V 1.002 ASK+ K X 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Warm-up time: 30 min. - REQUIRED 1.004 DISP Temperature: 18 - 28 degC 1.004 DISP Humidity: <80% 1.005 HEAD INITIAL SETUP 1.006 DISP While pressing the HOLD key, press POWER 1.006 DISP to turn UUT on. If Battery indicator illuminates, 1.006 DISP replace battery before proceeding. 1.007 DISP Remove the bottom cover and select K Type thermocouple. 1.008 DISP Connect the 5520A TC terminals to the CH A Input using 1.008 DISP K Type thermocouple wire and connectors. 1.009 DISP Press "A" on the front panel to select CH A input. 1.009 DISP Select AUTO range. 1.009 DISP Select degF. 1.010 HEAD {K TYPE} 1.011 5520 -250.0degF 0.1% 1.3U _K TC 2W 2.001 5520 32.0degF 0.1% 1.3U _K TC 2W 3.001 5520 2000degF 0.2% 1.8U _K TC 2W 4.001 DISP Select degC mode. 4.002 5520 -150.0degC 0.2% 1U _K TC 2W 5.001 5520 0.0degC 0.1% 0.7U _K TC 2W 6.001 5520 1370degC 0.2% 1U _K TC 2W 7.001 HEAD {J TYPE} 7.002 DISP Connect the 5520A TC terminals to the CH A Input using 7.002 DISP J Type thermocouple wire and connectors. 7.003 DISP On the HH82, select J Type thermocouple. 7.004 5520 -270.0degF 0.1% 1.3U _J TC 2W 8.001 5520 32.0degF 0.1% 1.3U _J TC 2W 9.001 5520 1800degF 0.2% 1.8U _J TC 2W 10.001 DISP Select degC. 10.002 5520 -160.0degC 0.2% 1U _J TC 2W 11.001 5520 0.0degC 0.1% 0.7U _J TC 2W 12.001 5520 990degC 0.2% 1U _J TC 2W 13.001 HEAD {HOLD FUNCTION} 13.002 5520 -100.0degC _J TC S 2W 13.003 DISP Press the HOLD key. 13.004 5520 * S 13.005 EVAL Does the {HOLD display = ~ 0.0degC}? 14.001 DISP Press the HOLD key to take out of HOLD. 14.002 TARGET 14.003 HEAD {REL FUNCTION} 14.004 5520 -100.0degC _J TC S 2W 14.005 PORT [D3000] 14.006 DISP Press the REL key. 14.007 5520 100.0degC _J TC S 2W 14.008 EVAL Does the {REL display = ~200degC}? 15.001 HEAD {MAX/MIN FUNCTION} 15.002 DISP Press the RCD (max min) key. 15.003 5520 -100.0degC _J TC S 2W 15.004 PORT [D3000] 15.005 5520 100.0degC _J TC S 2W 15.006 DISP Press the READ key. 15.007 EVAL Does the {READ display = ~100degC}? 16.001 DISP Press the READ key. 16.002 EVAL Does the {READ display = ~-100degC}? 17.001 5520 * S 17.002 HEAD {B INPUT} 17.003 DISP Connect the 5520A TC terminals to the CH B Input using 17.003 DISP J Type thermocouple wire and connectors. 17.004 DISP Press "B" on the front panel to select CH B input. 17.004 DISP Select AUTO range. 17.004 DISP Select degC. 17.005 5520 0.0degC 0.1% 0.7U _J TC 2W 18.001 END